lf_mask -- Verify operation of the CRaTER internal calibration signal
lf_ecal
none
This program connects to the CRaTER GSE with bcmd and CData 121 and, exercises the internal electrical calibration pulse generator in steps of 8 counts from an amplitude of zero to 248. This is done for both the low and high calibrate source input.
The output data is tagged with an ``echo'' preceding every amplitude change, 0x0A00 -> 0x0AF8 for the low range, 0x0B00 -> 0x0BF8 for the high range.
This script is part of the Long Form Functional Test. It assumes no particular instrument state and leaves the instrument in RESET.
0x8000 -> 0x80F8 preceding each low level cal input
0x8100 -> 0x81F8 preceding each high level cal input
0xFFFF -- end of data run
perl5.002 Minimum version of Perl interpreter required
CRATERTOOLS Environment variable containing path to CRaTER utilities
CRATER_GSE Environment variable containing EGSE machine name
Interpretation of the resulting data is not automated.
bcmd
Bob Goeke
$Id: lf_ecal,v 1.2 2007/09/13 14:51:57 goeke Exp goeke $